Initialization.ini
186 lines
| 7.2 KiB
| text/x-ini
|
IniLexer
/ PositioningSystem [LV2021] / Initialization.ini
| r0 | [Tester] | ||
| TesterType = "EHF1" | |||
| TesterName = "Elektrischer Hochfrequenzmessplatz 1" | |||
| [cHardware] | |||
| POS_AxesNames.<size(s)> = "3" | |||
| POS_AxesNames 0 = "X" | |||
| POS_AxesNames 1 = "Y" | |||
| POS_AxesNames 2 = "Z" | |||
| r5 | POS_Reference_Axes = "<size(s)=3> 200,000000 200,000000 1,000000" | ||
| POS_Limits_Axis1 = "<size(s)=2> 1 210" | |||
| POS_Limits_Axis2 = "<size(s)=2> 1 240" | |||
| POS_Limits_Axis3 = "<size(s)=2> 1 111" | |||
| r0 | POS_AxesFlipped.<size(s)> = "3" | ||
| POS_AxesFlipped 0 = "False" | |||
| POS_AxesFlipped 1 = "False" | |||
| POS_AxesFlipped 2 = "True" | |||
| r5 | Pos_MaxVelocity = "300" | ||
| Pos_MaxVelocityUnblockedDoor = "20" | |||
| Pos_MaxAcceleration = "15000" | |||
| Pos_MaxAccelerationUnblockedDoor = "20" | |||
| Pos_MaxJerk = "140000" | |||
| Pos_MaxJerkUnblockedDoor = "14000" | |||
| ForceLimit_Axis3 = "<size(s)=5> 0,000000 0,750000 0,780000 0,813000 0,920000" | |||
| ForceLimit_Axis3_Text.<size(s)> = "5" | |||
| ForceLimit_Axis3_Text 0 = "" | |||
| ForceLimit_Axis3_Text 1 = "�30 N" | |||
| ForceLimit_Axis3_Text 2 = "�40 N" | |||
| ForceLimit_Axis3_Text 3 = "�50 N" | |||
| ForceLimit_Axis3_Text 4 = "�60 N" | |||
| ForceLimit_Axis3_AO_Limits = "<size(s)=2> 0,000000 10,000000" | |||
| r0 | AO_ForceLimit = "Dev0/Ao0" | ||
| r5 | DI_E-Stop = "Dev0/AuxPort/Di0" | ||
| DI_Door = "Dev0/AuxPort/Di1" | |||
| r0 | |||
| [cEHF1] | |||
| POSConfig.IP = "10.1.2.1" | |||
| POSConfig.Port = "502" | |||
| DMMConfig.ResourceName = "COM1" | |||
| DMMConfig.Baudrate = "9600" | |||
| DMMConfig.ChannelList = "101,102,103,104,105,106,107,108,109,110" | |||
| r5 | ChargeMeterConfig.ResourceName = "COM10" | ||
| r0 | ChargeMeterConfig.Baudrate = "9600" | ||
| r4 | NetworkAnalyser.Names.<size(s)> = "2" | ||
| NetworkAnalyser.Names 0 = "Rohde & Schwarz (ZVA67)" | |||
| NetworkAnalyser.Names 1 = "Rohde & Schwarz (ZNL20)" | |||
| NetworkAnalyser.ResourceNames.<size(s)> = "2" | |||
| r5 | NetworkAnalyser.ResourceNames 0 = "TCPIP0::10.1.2.30::inst0::INSTR" | ||
| NetworkAnalyser.ResourceNames 1 = "TCPIP0::10.1.2.10::inst0::INSTR" | |||
| r4 | NetworkAnalyser.StateFilesFolder.<size(s)> = "2" | ||
| NetworkAnalyser.StateFilesFolder 0 = "C:\Rohde&Schwarz\Nwa\Calibration\RecallSets\" | |||
| NetworkAnalyser.StateFilesFolder 1 = "C:\Users\Public\Documents\Rohde-Schwarz\ZNL\Calibration\RecallSets\" | |||
| NetworkAnalyser.TestResultFolder.<size(s)> = "2" | |||
| r5 | NetworkAnalyser.TestResultFolder 0 = "D:\PHF4MeasurementsZVA67\EHF1\Tests\" | ||
| NetworkAnalyser.TestResultFolder 1 = "D:\PHF4MeasurementsZNL20\EHF1\Tests\" | |||
| r4 | NetworkAnalyser.TestResultFormat = "%d.%m.%Y-%H%M%S" | ||
| NetworkAnalyser.ResultFolder.<size(s)> = "2" | |||
| NetworkAnalyser.ResultFolder 0 = "D:\PHF4MeasurementsZVA67\" | |||
| NetworkAnalyser.ResultFolder 1 = "D:\PHF4MeasurementsZNL20\" | |||
| NetworkAnalyser.ResultFormat = "%d.%m.%Y-%H%M%S" | |||
| r0 | |||
| [cSCP] | |||
| DMMConfig.ResourceName = "COM1" | |||
| DMMConfig.Baudrate = "9600" | |||
| DMMConfig.ChannelList = "101,102,103,104,105,106,107,108,109,110" | |||
| ChargeMeterConfig.ResourceName = "COM2" | |||
| ChargeMeterConfig.Baudrate = "9600" | |||
| [cPathes] | |||
| pUser = "C:\Work\PositioningSystem\support\User.txt" | |||
| pResultsFolder = "C:\Work\PositioningSystem\support\ResultData" | |||
| r1 | pGraphs = "C:\Work\PositioningSystem\support\Force" | ||
| pExcelReports = "C:\Work\PositioningSystem\support\Force" | |||
| r6 | pTestRun = "C:\Work\PositioningSystem\support\TestRuns" | ||
| pMatrix = "C:\Work\PositioningSystem\support\Matrixes" | |||
| r0 | |||
| [cDefaultValues] | |||
| r6 | MovingParameter.Velocity = "20,000000" | ||
| MovingParameter.Acceleration = "50,000000" | |||
| MovingParameter.Jerk = "140000,000000" | |||
| r7 | ForceLimit_Z = "0" | ||
| r0 | Learning_Steps.X = "1,000000" | ||
| Learning_Steps.Y = "1,000000" | |||
| Learning_Steps.Z = "1,000000" | |||
| r7 | Learning_ZStopOffset = "2,000000" | ||
| r1 | Learning_Measurements = "3" | ||
| r4 | Learning_NetworkAnalyser = "0" | ||
| r0 | TestNumber = "1" | ||
| Comment = "" | |||
| Matrix_WorkingStroke = "4,000000" | |||
| Matrix_Z- = "1,000000" | |||
| Matrix_Contacts = "1" | |||
| Matrix_XDelta = "1,000000" | |||
| Matrix_XSize = "2" | |||
| Matrix_YDelta = "1,000000" | |||
| Matrix_YSize = "2" | |||
| Colors = "<size(s)=10> 16776960 16744448 16744640 16711935 10879142 8388863 8421631 423327 3381759 9567230" | |||
| r6 | TestRun_LoopsNumber = "2" | ||
| TestRun_Delay = "100" | |||
| TestRun_-Z = "1" | |||
| TestRun_Stroke = "4" | |||
| TestRun_Time = "0" | |||
| TestRun_StrokesNumber = "10" | |||
| r0 | |||
| r6 | |||
| r0 | [cReport] | ||
| FileNameFormat = "%s_TestResults.txt" | |||
| ReportNameFormat = "%s_TestReport.xlsx" | |||
| ReportTemplate_German = "C:\Work\PositioningSystem\support\ReportTemplates\TemplateGerman.xlsx" | |||
| ReportTemplate_English = "C:\Work\PositioningSystem\support\ReportTemplates\TemplateEnglish.xlsx" | |||
| r1 | ForceTemplate = "C:\Work\PositioningSystem\support\ReportTemplates\ForceTemplate.xlsx" | ||
| ResistanceForceDistanceTemplate = "C:\Work\PositioningSystem\support\ReportTemplates\ResistanceForceDistanceTemplate.xlsx" | |||
| r0 | |||
| [cStateMessages] | |||
| r5 | Info.<size(s)> = "28" | ||
| Info 0 = "Not-Halt aktiviert!" | |||
| Info 1 = "Einrichtbetrieb f�r Achsensteuerung aktiviert! Nur manuelle Positionierung m�glich" | |||
| Info 2 = "T�r ist freigegeben! Positionierung nur mit eingeschr�nkten Bewegungsparameter und bei bet�tigtem Zustimmtaster m�glich" | |||
| Info 3 = "Positioniereinheit wird initialisiert! Bitte warten..." | |||
| Info 4 = "Datenerfassungssystem wird initialisiert! Bitte warten..." | |||
| Info 5 = "Ladungsverst�rker wird initialisiert! Bitte warten..." | |||
| Info 6 = "Netzwerkanalysator wird initialisiert! Bitte warten..." | |||
| Info 7 = "Zum Anlegen eines neuen Tests "Neuer Test" klicken" | |||
| Info 8 = "Testdaten anlegen" | |||
| Info 9 = "Lernfahrten" | |||
| Info 10 = "Achsen werden initialisiert..." | |||
| Info 11 = "Referenzfahrt l�uft..." | |||
| Info 12 = "X-Achse wird referenziert..." | |||
| Info 13 = "Y-Achse wird referenziert..." | |||
| Info 14 = "Z-Achse wird referenziert..." | |||
| Info 15 = "Position wird angefahren..." | |||
| Info 16 = "Kraftmessung wird durchgef�hrt..." | |||
| Info 17 = "Widerstand/Kraftmessung wird durchgef�hrt..." | |||
| r7 | Info 18 = "Pr�flauf" | ||
| r5 | Info 19 = "Test geladen" | ||
| Info 20 = "Laufende Pr�fung..." | |||
| r7 | Info 21 = "Verz�gerung l�uft..." | ||
| Info 22 = "Messung wird durchgef�hrt..." | |||
| Info 23 = "Lastwechsel pro Position wird ausgef�hrt..." | |||
| Info 24 = "Pr�fung erfolgreich abgeschlossen" | |||
| Info 25 = "Fehler aufgetreten!" | |||
| Info 26 = "Test abgebrochen!" | |||
| Info 27 = "Testprotokoll wird erstellt... Bitte warten" | |||
| Info 28 = "Testprotokoll erstellt" | |||
| Result.<size(s)> = "29" | |||
| r0 | Result 0 = "Not-Halt" | ||
| Result 1 = "" | |||
| r5 | Result 2 = "" | ||
| Result 3 = "" | |||
| Result 4 = "" | |||
| r4 | Result 5 = "" | ||
| r5 | Result 6 = "" | ||
| Result 7 = "" | |||
| Result 8 = "Bereit" | |||
| r0 | Result 9 = "Einrichtbetrieb" | ||
| Result 10 = "Einrichtbetrieb" | |||
| Result 11 = "Einrichtbetrieb" | |||
| Result 12 = "Einrichtbetrieb" | |||
| r4 | Result 13 = "Einrichtbetrieb" | ||
| r5 | Result 14 = "Einrichtbetrieb" | ||
| Result 15 = "Einrichtbetrieb" | |||
| Result 16 = "Einrichtbetrieb" | |||
| Result 17 = "Einrichtbetrieb" | |||
| Result 18 = "Bereit" | |||
| Result 19 = "Bereit" | |||
| Result 20 = "Pr�fung l�uft" | |||
| Result 21 = "Pr�fung l�uft" | |||
| Result 22 = "Pr�fung l�uft" | |||
| r7 | Result 23 = "Pr�fung l�uft" | ||
| Result 24 = "Pr�fung fertig" | |||
| Result 25 = "Fehler" | |||
| Result 26 = "Pr�fung Stop" | |||
| r5 | Result 27 = "" | ||
| r7 | Result 28 = "" | ||
| r0 | |||
| [cDialogMessages] | |||
| r6 | Messages.<size(s)> = "10" | ||
| Messages 0 = "W�hlen Sie eine Konfigurationsdatei zum Laden aus" | |||
| Messages 1 = "W�hlen Sie einen Ordner aus, der die Testergebnisse zum Laden enth�lt" | |||
| r0 | Messages 2 = "Die Merkzetteldaten werden gel�scht!\0ABitte best�tigen Sie die Operation oder schlie�en Sie das Dialogfeld, um den Vorgang abzubrechen ..." | ||
| r6 | Messages 3 = "W�hlen Sie einen Pr�fablauf zum Laden aus" | ||
| Messages 4 = "Geben Sie einen Namen f�r den Pr�fablauf ein" | |||
| r7 | Messages 5 = "Die Pr�fmatrix wird gel�scht!\0ABitte best�tigen Sie die Operation oder schlie�en Sie das Dialogfeld, um den Vorgang abzubrechen ..." | ||
| r6 | Messages 6 = "W�hlen Sie eine Pr�fmatrix zum Laden aus" | ||
| Messages 7 = "Geben Sie einen Namen f�r die Pr�fmatrix ein" | |||
| Messages 8 = "W�hlen Sie eine Vorlage aus:" | |||
| Messages 9 = "Test wurde abgebrochen.\0AOption zur Achseninitialisierung w�hlen oder abbrechen." |